/script>

BACTERIAL ISOLATES ASSOCIATED WITH HOSPITAL ACQUIRED INFECTIONS FROM THERMOMETERS, STETHOSCOPES AND SPHYGMOMANOMETER CUFFS



ABSTRACT
Medical equipments (stethoscopes, thermometers and sphygmomanometer cuffs) which are universally used as a medical device by health workers, are likely to be contaminated by microorganisms, if they are not cleaned or disinfected may transmit pathogens from one patient to another. A total of 30 samples were collected from Irrua Specialist Teaching Hospital, Irrua, cultured and incubated at 370c for 24 hours to observe for growth. Fourteen (46.7%) of the medical equipments had bacterial contamination and sixteen  (53.3%) have no significant growth. Staphylococcus aureus was the highest contaminant with nine  isolates (64.3%) followed by Staphylococcus epidermidis with three isolates (21.4%) and the least isolated was Staphylococcus saprophyticus with one isolate (7.1%). Different biochemical tests were done to identify the isolates. Susceptibility to antibiotics of the isolates was also performed. Integration of medical equipments care in daily hospital routine enhances the control of nocosomial infections.

TABLE OF CONTENTS
Title page                                                                 i
Certification                                                             iii
Dedication                                                               iv
Acknowledgement                                                    v
Table of content                                                       vi
Abstract
CHAPTER ONE
1.1      Introduction                                                     1
1.2      Objective of  the study                                     4
CHAPTER TWO
2.1      Literature review.                                             5
2.2      Epidemiology                                                   6
2.3      Cause                                                              6
2.4      System                                                             8
2.5      Diagnosis                                                         9
2.6      Treatment                                                        11
2.7      Transmission                 
2.8   Risk Factors                                                    13
2.9 Prevention                                                         16
CHAPTER THREE
3.1      Study site                                                          19
3.2      Materials and method                                        19
3.3      Cleaning and sterilization                                  20
3.4      Preparation for sterilization                             20
3.5      Specimen                                                         20
3.6      Identification of Isolates                                   21
3.7      Biochemical test                                              22
CHAPTER FOUR
Result                                                                      25
CHAPTER FIVE
5.1      Discussion                                                       32
5.2      Conclusion and Recommendation                   35
Reference                                                                 38
Appendix 1                                                              47
Appendix 11                                                            50
CHAPTER ONE
1.1 INTRODUCTION
        Hospital acquired infections, also known as nocosomial infections, are infections whose development is favoured by the hospital environment, such as one acquired by a patient during the hospital visit, or one developing among staff. Infections are considered nosocomial when they become clinically evident during hospitalization (at least 72 hours after admission (Orrett et al., 1998). In developing countries, between 5% and 10% of patients acquire one or more infections, and 15-40% of patients admitted to critical care are thought to be affected (Lazzari et al., 2004; Klevens et al, 2007). In poor settings, such as most developing countries including Nigeria, the rate of infection can exceed 20% (Pittet, 2005; WHO, 2008). Hospital acquired infections exact a tremendous toll, resulting in increased morbidity and mortality, and increased health care costs (Haley, 1985). Infection transmission in the hospital environment remains a significant hazard for hospitalized patients and healthcare workers are potentially source of these infections, with many pathogens transmitted by medical device such as thermometers, sphygmomanometer cuffs and stethoscopes (Patent storm, 2004). These medical equipments are used in hospitals by medical doctors, medical students and other health practitioners for assessing patients’ health, have been reported as a potential formites for transmitting infections in the hospital environment in various parts of the world (Cohen et al., 1996; Zuliani-Maluf et al., 2002).
        There are increasing reports of the tremendous risk of transmitting antibiotic-resistant bacteria from one patient to another from thermometer, sphygmomanometer cuff and stethoscopes. Because most hospital acquired infections are primarily nosocomial and not auto infections (Hoog Kampkorstanze et al., 1982), their acquisition in the hospital environment adds to morbidity, mortality and economic costs (Parmar et al., 2002). Stethoscopes, thermometers and sphygmomanometer cuffs are the universal tools of the medical profession, are additional possible carriers, as these come in contact with many patients, following their contact with the skin; microorganisms can attach and establish themselves on the medical equipments and subsequently be transferred to other patients if the medical equipment is not disinfected before reuse (Sander, 2003; Madar et al., 2005). The transmission of infection from contaminated medical devices can be a possible cause of hospital acquired infection (nocosomial infection).
This accounts for some of the difference in the knowledge, attitudes and practices among medical professional in the use of the medical equipments (Stethoscopes, thermometers and sphygmomanometer cuffs).
1.2      OBJECTIVES OF THE  RESEARCH
1.     To assess medical equipments (Stethoscopes, thermometers and sphygmomanometer cuffs) handling and maintenance practices among Medical practitioners.
2.     To determine the bacterial agents that can contaminate medical equipments (Stethoscopes, thermometers and sphygmomanometer cuffs) used by medical practitioners.
3.     To determine the antibiotic sensitivity of bacterial isolates from the medical equipments used by
         the Medcal  Practitioners. 
CHAPTER TWO
2.1      LITERATURE REVIEW
The majority of Nosocomial infections are of non minor or moderate clinical importance but nevertheless may cause distressing morbidity, lengthen hospital stay and increase costs. Some nosocomial infections can have catastrophic local consequences, such as infection of a hip joint prosthesis or a prosthetic hearth valve, although these do not necessarily result in death. Serious nosocomial infections due to Gram negative bacilli and methicillin-resistant staphylococci have increased greatly in incidence during the last 30 years and the ultimate consequence of these may be fatal septicaemia (Cohen et al., 1997; Zuliani-Maluf et al., 2002).New or previously unrecognized serious problems have occasionally occurred in recent years, such as nosocomial infection caused by Vancomycin-resistant enterococci. Members of the hospital staff are also sometime at risk from other disease such as tuberculosis, hepatitis, chicken pox and human immuno deficiency virus (HIV) infections.
        There are three main routes of spread of organisms causing nosocomial infections: Self infection (autogenous infection), Gross-infection and environmental infection. Colonization rather than infection occurs with each of these three routes in many instance. This is particularly relevant when considering the epidemiology of outbreak of nosocomial infection, as the colonized patient or member of staff may cause symptomatic infection in another patient. Examples of nosocomial infection are : Ventilator associated pneumonia (VAP), Staphylococcus aureus infection, Methicillin resistant Staphylococcus aureus (MRSA) wound infections, Candida albicans infections, Pseudomans aeruginosa infections, Stenotrophomonos maltophilig infections, Pseudomenbraneous colitis, Tuberculosis, Urinary tract infection, Hospital acquired Pneumonia (HAP), Gastroenteristics, Vancomyan-resistant Enterococcus (VRE), and Legionanaire’s  disease.
2.2      EPIDEMIOLOGY
Nosocomial infections are commonly transmitted when hospital officials become complacent and personnel do not practice correct hygiene regularly. Increased use of inpatients treatment mean that people who are hospitalized are more ill and have more weakened Immune systems than may have been true in the past. Moreover, some medical procedure by pass the body’s natural protective barriers. Since medical staffs move from patient to patient, the staffs themselves serve as a means for spreading pathogens. Essentially the staffs and some medical equipment can serve as vectors.
2.3      CAUSES
All hospitalized patients are at risk of acquiring an infection from their treatment or surgery. Some patients are at greater risk than others, especially young children, the elderly and persons with compromised immune systems. The National Nosocomial Infection Surveillance System Database compiled by CDC shows that the overall infection among the children in intensive care is 6.19%, with the primary causes being venous catheters and ventilator-associated pneumonia (Patent storm, 2004).
The risk factors for hospital acquired infections in children include parenteral nutrition (tube or intravenous feeding), the use of antibiotics for more than 10 days, use of invasive devices, poor post-operative status and immune system dysfunction. Other risk factors that increase the opportunity for hospitalized adults and children to acquire infection are:
1.     A Prolonged hospitalization
2.     Severity of underlying illness
3.     Compromised nutritional or immune status
4.     Use of indwelling catheters
5.     Failure of health care workers to wash their hands between patients or before procedures
6.     Prevalence of antibiotic resistance bacteria from the overuse of antibiotics. Any type of invasive procedure can expose a patient to the possibility of infection. Some common procedures that increase the risk of hospital acquired infections include
1.     Urinary bladder catheterization
2.     Respiratory procedures such as incubation or mechanical ventilation
3.     Surgery and the dressing or drainage of surgical wounds
4.     Intravenous procedures for delivery of medication, transfusion or nutrition
5.     Gastric drainage tubes into the stomach through the nose or mouth.
2.4      SYMPTOMS
Fever is often the first sign of infection (Vincent, 2003). Other symptoms and signs of infection are rapid breathing, mental confusion, low blood pressure and a high white blood cell count. Patients with a Urinary Tract Infection (UTI) may have pain when urinating and blood in the urine. Symptoms of pneumonia may include difficulty breathing and cough  (Cohen et al., 1997; Zuliani-Maluf et al., 2002). A localized infection begins with swelling, redness and tenderness on the skin or around a surgical wound or other open wound, which can progress rapidly to the destruction of deeper layers of muscle tissue and eventually sepsis.
2.5      DIAGNOSIS
An infection is suspected anytime a hospitalized patient develops a fever that cannot be explained by the underlying illness. Some patients, especially the elderly, may not develop a fever. In these patients, the first signs of infection may be rapid breathing or mental confusion (Saxena et al., 2005).
        Diagnosis of a hospital acquired infection is determined by:
1.     Evaluation of symptoms and signs of infection.
2.     Examination of wounds and catheter entry sites for redness, swelling or the presence of pus or an abscess.
3.     A complete physical examination and review of underlying illness.
4.     Laboratory tests, including complete blood count (CBC) especially to look for an increase in infection fighting white cell, urinalysis, looking out for white cells or evident of blood in urinary tract; culture of infected area, blood, sputum, urine, or other body fluids to find the causative organism.
5.     Chest X-ray may be done when pneumonia is suspected to look for the presence of white blood cells and other inflammatory substances in lung tissue.
6.     Review of all procedure performed that might have led to infection.
2.6      TREATMENT
Cultures of blood, urine, sputum, other body fluids, or tissue are important in order to identify the bacteria, fungi, virus or other microorganism have being identified, it will be tested again for sensitivity to a range of antibiotics so that patients can be treated immediately and effectively with an appropriate medicine to which the causative organism will respond to (Bryan 1986; Pitte et al., 2008). While waiting for a test result, treatment may begin with common broad spectrum antibiotics such as penicillin, cephalosporins, tetracyclines or erythromycins. Some types of bacteria are becoming resistant to these standard antibiotic treatments, especially when patients with chronic illness are given antibiotic therapy for a long period of time when this happens, a different, more powerful, and more specific antibiotic must be used to which the specific organism responds to (Bryan, 1986; Pitte et al., 2000). Two strong antibiotics that have been effective against resistant bacteria are Vancomycin and Imipenem, although some bacteria are developing resistance to these drugs as well. The prolonged use of antibiotics is also known to reduce the effectiveness of the patient’s own immune system, sometimes becoming a factor in the development of infection. Fungal infections are treated with antifungal medications. Examples of these are amphotericin B, Nystatin, Ketoconazole, Itraconazole and Fluconazoles.

2.7   TRANSMISSION
The transmission of hospital acquired infections are shown below with their description:

Route
Description
1
Contact Transmission
The most important and frequent mode of transmission of hospital acquired infection (Nosocomial)
2
Droplet transmission
Occurs when droplets are generated from the source person mainly during coughing, sneezing and talking; and during the performance of certain procedures such as bronchoscopy. Transmission occurs when droplets containing microorganisms from the infected person are propelled through air and deposited on the host body.
3
Air Borne Transmission
Occurs by dissemination of either airborne droplet nuclei (small particle residue, 5mm or smaller in size) of evaporated droplets containing micro organism suspended in air for a long period or dust particles containing the infectious agents. Microorganisms carried in this manner can be dispersed widely by air currents and may be inhaled by a susceptible host within the same room or over a long distance from the source patients, depending on the environmental factors. Therefore, special air handling and ventilation are required to prevent air-borne transmission. Microorganism transmitted by air include Legionella, Mycobacterium tuberculosis and Rubeola.
4
Common vehicle transmission
Applies to microorganisms transmitted to the host by contaminated items such as food, water, medications, devices and equipment.
5
Vector borne transmission
Occurs when vectors such as mosquitoes, flies, rats and their vermin transmit microorganisms.

Contact transmission is divided into two subgroups which are:
i.      Direct contact transmission and
ii.     Indirect contact transmission
Rates of contact Transmission

Routes
Description
1
Direct contact transmission
Involves a direct body surface contact and physical transfer of microorganisms between a susceptible host and an infected or colonised person turns a patient, gives a patient a bath or performs other patient care activities that require direct personal contact. Direct contact transmission also occurs between two patients, with one serving as the source of the infections.
2
Indirect transmission
Involves a contact of a susceptible host with a contaminated intermediate object, usually inanimate such as contaminated instrument or needle. Gloves that are changed between patients. In addition, the improper use of saline flush syringes, vials, and bags has been implicated in disease transmission. Even when Health workers had access to gloves, disposable needles, intravenous device and flush.
  2.8  RISK FACTORS
Factors predisposing a patient to infection can broadly be divided in three areas:
1.     People in hospitals are usually in a poor state of health, imparing their defences against bacteria. Advanced age or premature birth along with immunodeficiency (due to drugs, illness or irradiation) present a general risk while other diseases can present specific risk for instance, chronic obstructive pulmonary disease can increase chances of respiratory tract infection (Parmar et al., 2004; Lange et al., 2000).
2.     Invasive devices, for instance incubation tubes, catheters, surgical drains and tracheostomy tubes all bypass the body’s natural line of defence against pathogens and provide an easy route for infection. Patients already colonized on admission are instantly put at greater risk when they undergo an invasive procedure (WHO, 2003).
3.     A patient’s treatment itself can leave them vulnerable to infection. Immuno-suppression and antacid treatment undermine the body’s defence, while antimicrobial therapy (removing competitive flora and leaving resistant organisms) and recurrent blood transfusion have also be identified as risk factors.
2.9   PREVENTION
Hospital acquired infections can be prevented through the various steps:
1.       Adoption of an infection control program such as the one sponsored by U. S. Centres For Diseases Control (CDC), which includes quality of procedure known to lead infection and monitoring program to track infection rate to see if they go up or down.
2.       Sterilization of all reusable equipment such as thermometers, stethoscopes and sphygmomanometer cuffs and any other device that comes to contact with the skin.
3.       Strict attention to aseptic techniques in the performance of procedures including the use of gloves, nose masks and barriers.
4.       Reduction in the general use of antibiotics to encourage better immune response in patients and reduce the cultivation of resistance.
5.       Isolation of patients with known contagious infection
6.       Frequent changing of dressing for wounds and use of antibacterial ointment.
7.       Limitation on the use and duration of high risk procedures such as urinary catheterization.
8.       Sterilization of medical equipment.
9.       Strict adherence to hand washing rule by Health care workers and visitors to avoid passing infectious microorganisms to or between hospitalized patients.
10.     Use of silver alloy coated urinary catheters that destroy bacteria before they can migrate to the bladder.

CHAPTER THREE
MATERIALS AND METHODS
3.1      STUDY SITE
The research was conducted in Irrua Specialist Teaching Hospital (I.S.T.H) located at Esan Central Local Government, Edo State, Nigeria. The study entails the bacterial isolates associated with medical equipments from thermometers, stethoscopes and sphygmomanometer cuffs in various wards of the Teaching Hospital which include the male surgical wards (M.S.W.), Female surgical ward (F.S.W.), female medical wards (F.M.S), Ganaecology ward (G.W.) and from Pediatric wards. The research started on December 1 and ended December 15, 2011.
3.2      MATERIALS AND EQUIPMENT USED:
Sterile Forceps, Petri dishes, sterile wire loops, incubator, Autoclave, Bunsen burner, Hot Air Oven, Gas, Distilled water, conical flask, cotton wool, marking pen, microscope, glass slides, cover slips, culture plates and Sensitivity disk.
3.3       CLEANING AND STERILIZATION OF GLASS WARE
All glass wares were cleaned with detergent water and rinsed with sterile distilled water before sterilization at 1600c for 1 hour in a hot air oven.
3.4      PRPARATION AND STERILIZATION OF MEDIA.
The media used for this study were blood agar, chocolate agar, MacConkey agar and Nutrient agar.
Media were reconstituted with water according to manufacturer’s instruction (See Appendix 1).
3.5      SPECIMEN
A total of 30 samples were collected from medical equipment such as stethoscopes, thermometers and sphygmomanometer cuffs from Irrua Specialist Teaching Hospital, Irrua. A sterile swab stick moistened in sterile physiological saline was swabbed all over the surface of each medical equipment and transferred for analysis to the medical microbiology laboratory of Irrua specialist teaching hospital, Irrua. All laboratory analysis were done within 1 hour of sample collection. The swabs were directly inoculated on blood agar, chocolate agar and MacConkey agar. The inoculated media were incubated for 24hours, 48 hours and 72 hours at 370c and then examined for bacterial growth (Cheesbrough, 2000) according to the standard protocol.
3.6      IDENTIFICATION OF ISOLATES
GRAM STAINING: This technique helps to group organisms into Gram positive and Gram negative. It also shows the shapes of the organisms (rod, spiral or cocci) and the arrangement of the organism (in chain, clusters, singly) (Cheesbrough, 2000). Colonies of the organism were aseptically picked with a sterile wire loop and was then emulsified in a drop of normal saline on a slide. This was allow to air dry and then fixed with heat by passing through heat briefly (Cheesbrough, 2000).
        The smear was stained first with crystal violet solution for 30 seconds, then washed off under running water. Lugol’s iodine solution was applied for 30 seconds. The stained smear was then decolourised with acetone and was immediately flooded off with running water. Then, stained using 1% neutral red for 30 seconds, after which it was flooded off with running water. The stained slide was allowed to air dry, after which a drop of oil immersion was put on the stained portion of the slide and the viewed under oil immersion objective lens (100x).
        All gram negative organisms were dark red colour while the organisms that stain dark purple is Gram positive.
3.7      BIOCHEMICAL TESTS
Biochemical tests were carried out for identification and characterization of various isolates.
        A pure culture of each of the isolates was used in these biochemical tests, they were obtained by sub culturing primary plate. The following tests were carried out for identification.
CATALASE TEST
        The catalase test method used was according to Harrigan and McCane (1976). An inoculum was taken from a pure culture of the organism which was emulsified in
 normal saline on a clean grease-free slide. A drop of hydrogen peroxide (H2O2) was added into the suspension using a Pasteur pipette. The slide is observed carefully for gas bubbles, if present, it indicates a positive test while the absence of gas bubbles indicates a negative test.
COAGULASE TEST
        An inoculum was taken from pure culture of the organism with the aid of a sterile wire loop and emulsified in normal saline on a grease-free slide, a drop of plasma was added and the slide was gently rocked. A positive result is represented by the coagulation of the plasma indicating the production of coagulase. This was done to confirm the presence of Staphylococcus aureus in Gram positive cocci organisms and those which were catalase positive. The enzyme, coagulase produced by Staphylococcus aureus converts fibrinogen in plasma to fibrin forming agglutinates
SUGAR FERMENATION TEST
        The test was carried out following Harrigan and McCane (1976) procedure. One percent (1%) solution of a series of sugars were prepared in peptone water to which neutral red indicator was added. This was then distributed equally into bijou bottles and sterilized at 1210C for 10minutes after cooling. A loopful of the pure colony of each test organism which were grown for three to four hours in sterile peptone water at 370C was then inoculated into sterile 1% sugar solution in peptone water. The set was then incubated at 370C for 24 hours. The sugars used were Mannitol and Sucrose. The production of a yellow to pink colour indicated acid production while the trapping of gas in an inverted Durham’s tube indicated gas production.
CHAPTER FOUR
RESULTS
A total of 30 samples were taken from medical equipment (stethoscopes, thermometers and sphygmomanometer cuffs) from Irrua specialist teaching hospital, Irrua in various wards of the hospital - the female medical ward, male surgical ward, Gynaecology ward, Female surgical ward and paediatric ward. Fourteen (46.7%) of the medical equipment had bacterial contamination and sixteen (53.3%) had no significant growth after 72 hours.
The table below show the growth of bacteria after incubating for 24 hour, 42 hours and 72 hours at room temperature. 
Table 1:    Occurrence of microorganisms   in the medical equipments                                                         
                                                          Hours of Incubation
Ward
Material
24
48
72
F.M.W
Stethoscope 1           Diaphragm
-
-
-

                                       Bell
-
-
-

Stethoscope 2            Diaphragm
-
+


                                       Bell
-
+


Thermometer
-
+


Sphygmomanometer cuff 1
-
+


Sphygmomanometer cuff 2
+



Sphygmomanometer cuff 3
+


M.S.W
Sthethoscope             Diaphragm
-
-
-

                                        Bell
-
-
-

Sphygmomanometer cuff
-
-
+
G.ward
Thermometer 1
-
-
-

Thermometer 2
-
-
-

Stethoscope               Diaphragm
-
-
-

                                        Bell
-
-
-

Sphygmomanometer Cuff
+


F.S.W
Stethoscope 2             Diaphragm
-
-
-

                                         Bell
-
-
-

Stethoscope 2              Diaphragm
-
+


                                          Bell
-
-
-

Sphygmomanometer cuff 1
-
+


Sphygmomanometer cuff 2
-
+


Sphygmomanometer cuff 3
-
+

P.ward
Stethoscope 1                Diaphragm
-
-
-

                                         Bell
+
+


Stethoscope                    Diaphragm
-
-
-

                                            Bell
-
-
-

Sphygmomanometer cuff
+



Stethoscope                     Diaphragm
-
-
-

                                              Bell
-
-
-

KEY:
+ = growth
-      = no growth
F.M.W = Female Medical Ward
P. Ward = Paediatric ward
F.S.W.  = Female Surgical Ward
M.S.W.  = Male Surgical Ward
Table 2:
Ward, material and the bacterial isolated
Ward
Material
Isolate
F.M.W
Stethoscope Diaphragm
Staphylococcus Saprophyticus

                       Bell
Staphylococcus aureus

Thermometer
Staphylococcus aureus

Sphygmomanometer cuff 1
Staphylococcus aureus

Sphygmomanometer cuff 2
Staphylococcus epidermidis

Sphygmomanometer cuff 3
Staphylococcus epidermidis
M.S.W.
Sphygmomanometer cuff
Staphylococcus aureus
G.ward
Sphygmomanometer cuff
Staphylococcus aureus
F.S.W
Stethoscope 2 Diaphragm
Staphylococcus aureus

Sphygmomanometer cuff 1
Staphylococcus aureus

Sphygmomanometer cuff 2
Staphylococcus aureus

Sphygmomanometer cuff 3
Staphylococcus aureus
P.ward
Stethoscope Bell
Staphylococcus epidermidis

Sphygmomanometer cuff
Staphylococcus aureus

KEY:
F.M.W. = Female medical ward,
M.S.W = Male Surgical ward,
P. Ward = Paediatric ward
G. Ward = Gynaecology ward
F.S.W = Female Surgical Ward.
Table 3: Bacterial Isolates from medical equipment with their number of isolates and their percentages
Bacterial isolate
No isolated
% isolated
Staphylococcus aureus
9
64.3%
Staphylococcus epidermidis
3
21.4%
Staphylococcus saprophyticus
1
7.1%
Total
14
46.7%

From the organisms isolated in Irrua specialist teaching hospital, Irrua, Staphylococcus aureus was the most frequent organism with nine isolates (63.4%) followed by Staphylococcus epidermidis  with three isolates (21.4%) and Staphylococcus saprophyticus  was the least organism to be isolated with one isolate (7.1%).
Table 4: Biochemical characteristics of bacterial isolate
Characterization
S. aureus
S. epidermidis
S. saprophyticus
Gram reaction
+
+
+
Catalase
+
+
+
Coagulase
+
-
-
Mannitol*
+
-
-
Sucrose*
+
+
+
KEY:
* = Fermentation test
+ = Positive
- = Negative
        Susceptibility pattern to some antimicrobial agents is shown in table 5 below. Isolates of Staphylococcus aureus were more sensitive to Cefixime, Ciprofloxacin and Gentamycin (60%), Staphylococcus epidermidis were more sensitive to Ciprofloxacin, Gentamycin, Graxine (100%) and Staphylococcus saprophyticus Cefixime, were more sensitive to Ciprofloxacin, Gentamycin and Graxine (100%).
Table 5: Susceptibility patterns of isolates.
                                       No. of Sensitive isolate (%)
Antibiotics
Abbreviation
Staphylococcus aureus
Staphylococcus epidermidis
Staphycoccus Saprophyticus
Erythromycin
E
4 (40%)
0 (0)
0 (0)
Cefixine
CEF
6 (60%)
0 (0)
1 (100%)
Ciprofloxacin
CIP
6 (60%)
3 (100%)
1( 100%)
Gentamycin
GN
6 (60%)
3 (100%)
1 (100%)
Augumentine
AU
0 (0)
1 (33.3%)
0( 0)
Cotrimoxazole
COT
0 (0)
0 (0)
0( 0)
Cloxaculine
CL
0 (0)
0 (0)
0( 0)
Tetracyhile
TET
0 (0)
0 (0)
0 (0)
Graxine
GR
2 (20%)
2 (66.7%)
1 (100%)
Ampicillin
AM
0 (0)
0 (0)
0(0)

Number of Isolated Staphylococcus aureus = 10
Number of Isolated Staphylococcus epidermidis =3
Number of Isolated Staphylococcus Saprophyticus=1
CHAPTER FIVE
DISCUSSION, CONCLUSION & RECOMMENDATION
5.1   DISCUSSION
The rate of contamination (46.7%) observed in this study indicates that the medical equipment used by health care worker’s could be vectors playing a major role in transmitting microorganisms in the hospital environment. Earlier studies have also indicated that insufficient emphasis on consistent cleaning of medical equipment (Stethoscopes, Thermometers and Sphygmomanometer cuffs) in the medical curriculum are responsible for the high rate of bacterial contamination of the health workers (Zuliani-Maluf et al., 2002; Osorib et al., 2000). A number of studies have demonstrated that 47-60% of medical equipment analysed were colonized by various species of bacterial agents (Cohen et al., 1997, Smith et al., 1996, Sander, 2003, Madar et al., 2008).
        The spectrum of organisms isolated in this study was also isolated in a number of previous studies (Zuliani-Maluf et al., 2000l Sander, 2003; Madar et al., 2005). Of the bacteria isolated from stethoscopes, thermometers and sphygmomanometer cuffs in this study, Staphylococcus aureus was the most commonly isolated contaminant with 9 (64.3%), followed by Staphylococcus epidermidis 3 (2.49) and the least isolated Staphylococcus saprophitycus 1 (7.1%).
Earlier studies showed that Staphylococcus has developed resistance to conventional antibiotics (WHO, 2000; Sood et al., 2000) and the findings of this research confirm this.      The antibiotic sensitivity test conducted in our study indicated that all the isolated bacteria showed high levels of resistance to most of the antibiotics assessed (WHO, 200; Lange et al., 2000; Uneke and Ogbu, 2007). It is well proven that these antibiotic resistant microorganisms are capable of initiating severe nosocomiasis in a hospital environment and could require contact isolation and aggressive treatment to prevent their spread (Gupta et al., 2004; Saxena et al., 2005; Breathnach et al., 1992).
This study reinforces the dire need to revisit the medical curriculum with view to integrating adequate cleaning of thermometers, stethoscopes and sphygmomanometer cuffs, as a strategy of controlling nosocomial infections.
In most health care settings, the prevention of nosocomial infections is given serious consideration. Unfortunately however, primary attention to preventing nosocomial infections is usually passed through high risk invasive diagnostic tools and therapeutic health care procedures.
        The importance of the simple procedures such as hand hygiene and sterilisation of health care tools (stethoscopes, thermometers and sphygmomanometer cuffs) have been underestimated (Sengupta et al., 2006; Madar et al., 2009). By virtue of constant contact with patients by touch and by their medical devices, health care workers have become a potential source of transmission of hospital acquired infections. All need to wash their hands before and after seeing each patient. Failure to do so could facilitate the introduction of pathogens on any device that the health worker uses frequently, such as the thermometers, sphygmomanometer cuffs and stethoscopes. An earlier study showed that bacterial colony counts were significantly reduced from the stethoscopes, thermometers and sphygmomanometer cuffs after cleaning with isopropyl alcohol, sodium hypochlorite or benzalkonium chloride (Marinella et al., 1997).
5.2   CONCLUSION AND RECOMMENDATION
Hospital acquired infections represent an increasing financial burden and declining quality of health care in poor settings such as inmost developing countries including Nigeria and these have lead to increase in morbidity and mortality of the patient and health care workers. The need of strict prevention guidelines is very essential (Bernard et al., 199; Harris et al., 2000).
        One possible strategy for the prevention of hospital acquired infection outbreaks can be achieved by providing each patient with a disposable sphygmomanometer cuffs that will remain with them during their hospital stay and be disposed when the patients are discharged. Likewise, providing medical devices (thermometers and stethoscopes) in each patient room that are appropriately sanitized or disinfected between patients can prevent the outbreak of nocosmial infections. Strict adherence to CDC guidelines regarding hand washing, hand hygiene and use of standard precautions also remains critical to preventing future hospital acquired infections.
        Health professionals are used to carrying stethoscopes on the neck or in briefcases and they may take it home. The possibility of the transmission of the organisms from hospital to homes and vice versa, with the spread of microorganisms in the family members also needs to be explored and prevented (Lange et al., 2000; Klevens et al., 2002).
The need of frequent cleaning of stethoscopes, thermometers and sphygmomanometer cuffs with alcohol will reduce the transmission rate of nocosomial infections. Furthermore, it has been suggested that hospitals need to develop more rigorous programs and protocols for disinfecting their medical equipment such as thermometers, sphygmomanometer cuffs and stethoscopes (Myers et al., 1978; Jarvis, 2001).
REFERENCES

Africa-Purino, F. M. C., Dy, E. E. R. and Coronel, R. E. (2000), Stethoscopes: A potential source of Nosocomial Infections. Philippines journal Microbiology & Infection Diseases. 29(2): 9-13.
AOAC. (1990). Association of Official Analytical Chemists. Official Methods of Analysis 15th Edition (Helrick, K.ed.). AOAC, Arlington, Virginia.
Bernand, L., Lereveur, Durand, A. D., Goldstein, F., Mainardi, J. L., Acar, J. and Carlet, J. (1999). Bacterial Contamination of Hospital Physician’s Stethoscopes. Infection Control & Hospital Epidermiology. 2:626-8.
Breathnach, A. S., Jenkins, D. R., and Redler, S. J. (1992). Stethoscope as possible vectors of infected by Staphylococci. British medical Journal. 305: 1573-15744.
Bryan, C. S. (1986). Of Soap and Semmelweis. Infection Control. 7: 445-447.
Cheesbrough, M. (2000). District Laboratory Practice in Tropical Countries, Part 2, Cambridge University Press, UK, pp. 34-243.
Cohen, H. A. J., Amir, A., Matalon, R., Mayan, S., Beni and Barzilai, A. (1997). Stethoscopes and Otoscopes-A Potential Vector of Infection. Family Practice 14(6): 446-449.
Gastmeier, P. K., Groneberg, K., Weist and  Rüden, H. (2003). A cluster of Nosocomial Klebsiella pneumoniae Bloodstream infections in a Neonatal intensive Care Department: Identification of Transmission and Intervention. American Journal of Infection Control 31(7): 424-430.
Genné, D., de Torrenté, A., Humair, L., and Siegrist, H. H. (1996). Level of Stethoscope Contamination in the Hospital Environment. Shweizzerische Medizinische Wochenschrift. 126(51-52): 2237-40.
Gupta, A. P., Della-Latta, B., Todd, J., San Gabriel, Haas, F., Wu, D., Rubenstein, D. and Saiman. L. (2004). Outbreak of Extended-Spectrum Beta-Lactamase-Producing Klebsiella pneumoniae in a Neonatal Intensive Care Unit Linked to Artificial Nails. Infection Control & Hospital Epidemiology, 25(3): 210-215.
Haley, R. W., Culver, D. H., White, J. W., Morgan, W. M., Emori, T. G. and Munn, V. P. (1985). The Efficacy of Infection Surveillance and Control Programs in preventing Nosocomial Infection in US Hospitals. American Journal of Epidemiology. 121: 182-205.
Harris, A. D., Samore, M. H., Nafziger, R., Dirosario, K., Roghmann, M. C., and Carmeli. Y. (2000). A Survey on Handwashng Practices and Opinions of Healthcare Workers. Journal of Hospital Infection 45: 318-321.
Hoogkamp-Korstanje, J., Am, B., Cats, R., Senders C., and van Ertbruggen, I. (1982). Analysis of Bacterial Infections in a Neonatal Intensive Care Unit. Journal of Hospital Infection. 3: 275-284.
Jones, J. S., Hoerle, D. and Riekse, R. (1995). Stethoscopes: A potential Vector of Infection. Annals of Emergency Medicine. 26: 296-299.
Kerr, J. R., Martin, H., Chadwick, M. V., Edwards, A., Hodson, M. E. and Geddes, D. M. (2002). Evidence against Transmission of Pseudomonas aeruginosa by Hands and Stethoscopes in a Cystic Fibrosis Unit. Journal of Hospital Infection. 50(4): 324-326
Klevens, R., Edwards, J. R., Richards, C. L., Horan, R. C., Gaynes, R. P., Pollock, D. A. and Cardo, D. M. (2007). Estimating Health Care-Associated Infections and Deaths in U.S. Hospital. Public health Reports. 122: 160-166.
Lange, C. G., Morrissey, A. B. and Donskey, C. J. (2000). Point-Prevalence of Contamination of Healthcare workers’ Stethoscopes with Vancomycin-Resistant Enterococci at Two Teaching Hospitals in Cleveland, Ohio. Infection Control & Hospital Epidemiology. 21(12): 756.
Larson, E. (1988). A Causal Link Between Handwashing and Risk of Infection - Examination of the Evidence. Infection control and Hospital Epidemiology. 9: 28-36.
Lazzari, S., Allegranzi, B. and Concia, E. (2004). Making Hospitals Safer: the Need for a Global Strategy for Infection Control in Healthcare Settings. World Hospitals and Health Services, 32: 26-42.
Madar, R., Novakova, E. and Baska, T. (2005). The Role of Non-Critical Health-Care Tools in the Transmission of Nosocomial Infections. Bratislavske Lekarske Listy. 106(11): 348-350.
Marinella, M. A., Pierson, C. and Chenoweth, C. (1997). The Stethoscope: a Potential Source of Nosocomial Infection. Archives of Internal Medicine. 157: 786-790.
Orrett, F. A., Brooks, P. K. and Richardson, E. G. (1998). Nosocomial Infections in a Rural Regional hospital in a Developing Country: Infection Rates by Site, Service, Cost, and Infection Control Practices. Infection Control & Hospital Epidemiology. 19(2): 136-140.
Osorio, N. G., Oro, C., Ong, B. T., and, Emmanuel E. R. (2000). The Effect of a Traditional Educational Intervention on the Knowledge, Attitudes and Practices Regarding Stethoscope Use Among UST Senior Infectious Tropica Medicine Students. Philippines Journal of Microbiology & Infections Diseases. 29(4): 153-155.
Parmar, R. C., Valvi, C. C., Sira, P. and Kamat, J. R. (2004). A Prospective, Randomised, Double-Blind Study of Comparative Efficacy of Immediate versus Daily Cleaning of Stethoscope Using 66% Ethyl Alcohol. Indian Journal of Medical Sciences. 58(10): 423-430.
Patent S. US Patent Issued in 2004. Disposable Cover for Stethoscope Head. Retrieved June 15, 2008.
Pitter, D., Infection Control and Quality Health Care in the New Millenium. American Journal of Infection Control and Hospital Epidemiology. 21:381-386.
Pittet, D., Mourouga, P. and Perneger, T.V. (1999). Compliance with Handwashing in a Teaching Hospital. Annals of Internal Medicine. 130: 126-130.
Sanders, S. (2003). The Stethoscope and Cross-Infection. British Journal of General Practice. 53(497): 971-972.
Saxena, A. K., Panhotra, B. R. and Al-Mulhim, A. S. (2005). Contaminated Physician’s Stethoscope - A Potential Source of Transmission of Infection in the Hospital. Need of Frequent Disinfection after Use. Saudi medical Journal. 26(2): 348-350.
Sengupta, S., Sirkar, A. and Shivananda, P. G. (2000). Stethoscopes and Nosocomial Infection. Indian Journal of Pediatrics. 67(3): 197-199.
Smith, M. A., Mathewson, J. J., Ulert, A., Scerpella, E. G. and Ericsson, C. D. (1996). Contaminated Stethoscopes Revisited. Archives of Internal Medicine. 156: 82-84.
Sood, P., Mishra, B. and Mandal, A. (2000). Potential Infection Hazards of Stethoscopes. Journal of Indian Medial Association. 98(7): 368-370.
Tikhomirov, E. (1987). WHO Programme for the control of Hospital Infections. Chemiotherapia. 3: 148-151.
Uneke, C. K. and Ogbu, O. (2007). Potential for Parasite and Bacteria Transmission by Paper Currency in Nigeria. Journal of Environmental Health. 69(9): 54-60.
Vincent, J. L. (2003). Nosocomial Infections in Adult Intensive Care Units. Lancet. 361: 2068-2077.
World Health organization. (2008). 10 Facts on patients’ Safety.
World Health Organization, (2000). Overcoming Antimicrobial Resistance.
World Health organization. (2003). Manual for the Laboratory Identifications and Antimicrobial Sensitivity Testing of Bacterial Pathogens of Public Health Importance in Developing World WHO/DCS/CSR/RMD/2003, 6. Geneva. (pp 103-120).
Wright, I. M., Orr, H. and Potter, C. (1995). Stethoscope Contamination in the Neonatal Intensive Care Unit. Journal of Hospital Infection. 29(1): 65-68.
Zuliani-Maluf, M. E., Maldonado, A. F., Bercial, M. E. and Pedroso, S. A. (2002). Strethoscope: a Friend or an Enemy. Sao Paulo Medical Journal. 120(1): 13-15.
 APPENDIX I
PREPARATION OF MEDIA:
PREPARATION OF MACCONKEY AGAR
FORMULA                                          GRAMS/LITRE
Peptic digest of animal tissue                  20.00
Lactose                                                    10.00
Bile salts                                                  5.00
Sodium chloride                                      5.00
Neutral red                                               0.075
Final pH (at 25ºC)                                     7.4±0.2
DIRECTION/PREPARATION
        Suspend 550.7grams in 1000ml of distilled water. Heat to boiling, dissolve the medium completely, sterilize by autoclaving at 151bs pressure (121ºC) for 15 minutes. Avoid overheating. Cool to 45-50ºC and pour into sterile Petri dishes. The surface of the medium should be dry when inoculated.
PREPARATION OF NUTRIENT AGAR
FORMULA                                                       GRAM/LITRE
Peptic digest of animal tissue                          5.00
Beef extract                                                      1.50
Yeast extract                                                    5.00
Sodium chloride                                              5.00
Agar                                                                         15.00
Final pH (at 25ºC)                                            7.4±0.2
DIRECTION/PREPARATION
        Prepare by dissolving 28 grams of media in 1 litre of distilled water, heat to boiling, dissolve the medium completely. Sterilize by autoclaving at 121ºc, 15bs pressure for 15 minutes.
PREPARATION OF BLOOD AGAR
FORMULA                                                       GRAM/LITRE
Peptic digest of animal tissue                          5.00
Beef extract                                                      1.50
Yeast extract                                                    5.00
Sodium chloride                                              5.00
Agar                                                                15.00
Final pH (at 25ºC)                                            7.4±0.2
Sterile Blood
DIRECTION/PREPARATION
        Prepare by dissolving 28 grams of media in 1 litre of distilled water, heat to boiling, dissolve the medium completely. Sterilize by autoclaving at 121ºc, 15bs pressure for 15 minutes.
Add sterile blood and mixed thoroughly.
NUTRIENT AGAR
Composition
Ingredients                                       Grams/litre
Meat extract                                             1.0
Yeast extract                                            2.0
Peptone                                                    5.0
Sodium dilloride                                      5.0
Agar                                                         15.0
Final PH                                                   7.4 ± 0.2
Directions/Preparation
        28.0g nutrient base powder was weighed and dissolved in 1000ml of distilled water in a sterile conical flask. The medium was allowed to soak for 10 minutes swirled to mix, plugged with sterile cotton wool and then sterilized by autoclaving at 1210c for 15 minutes. It was allowed to cool to 470c, mixed well before pouring aseptically into sterile Petri-dish and allowed to set.
APPENDIX II
REAGENTS
CATALASE REAGENT:
Hydrogen Peroxide (H2O2) 3% of solution.
It is stored in a cool place with bottle closed and protected from light.
COMPOSITION OF GRAM STAINING REAGENTS:
PREPARATION OF CRYSTAL VIOLET SOLUTION
FORMULA                                               GRAMS/LITRE
Crystal violet                                                    2.0
95% Ethyl alcohol                                            20.0
Ammonium oxalate monohydrate                    0.2
Distilled water                                                  40.0
DIRECTION/PREPARATION
        The crystal violet was dissolved in 20ml of 95% ethyl alcohol and 0.2grams of ammonium oxalate monohydrate was dissolved in 20ml of distilled water. The solution were mixed and made with another 20ml of distilled water.

LUGOL’S IODINE SOLUTION
Lugol’s iodine 1g
Potassium ioding 2g
Distilled water 100ml
DIRECTION/PREPARATION
        The potassium iodine was first dissolve in the water, iodine crystal were then added and the mixture shorten properly to get completely dissolved.
PREPARATION OF SAFRANIN SOLUTION
FORMULAR                                             GRAM/LITRE
Safranin                                                           2.5
95% ethyl alcohol                                            100.0
Distilled water                                                  100.0
DIRECTION/PREPARATION
        The dye was dissolved in the alcohol, 10% of this solution was added to water (100ml) and stored thoroughly.

No comments:

Post a Comment